仕様

PERFORMANCE
Range C1 – 0.2 to 100%
C2 – 0.2 to 65%
C3 – 0.2 to 20%
nC4 & iC4 – 0.2 to 5%
nC5 & iC5 – 0.2 to 5%
LDD 2000 ppm short path cell
(Consult factory for lower LOD requirements)
Accuracy 5% of reading or 0.2% absolute, whichever is greater
Repeatability < 3% of reading or 0.2% absolute, whichever is greater
Response time 10 seconds
Flowrate 0.5 to 3.0 lpm, non-condensing, non-corrosive gas
ENVIRONMENTAL
Analyzer operational temperature range 10°C to 35°C
Field unit packaged operational temperature range -40°C to 50°C
Housing General purpose, purged NEMA, explosion proof, or pelican case
USER INTERFACE / CONNECTIVITY Modbus or SOAP over TCP/IP
HTML 5 User Interface
FORM FACTOR
Dimensions 9.5” Wide x 14” Long x 6” High (sensor unit without environmental packaging)
Weight 30 lbs (Pelican)
POWER REQUIREMENTS 100 to 240 VAC, 50/60 Hz, 200 Watts
DC power option available